Frequency range:
Standard features:
Absorber lining:
Typical Options:
Typical standards:
|
|
|
This is one of the original industry standard EMC Compact Anechoic Chambers designed for EMC full compliant radiated immunity testing for EN61000-4-3 and pre compliant radiated emission measurements for EN30147 part 2 /ANSI C63.4. For the 18 GHz option with hybrid absorber, the internal width of the chamber is increased to 3.3 m as shown in the photograph. |
Nominal size :
| 6.9 m long x 3 or 3.3 m wide x 3 m high |
Frequency range:
| Emissions 30 MHz to 1 GHz (option 18 GHz) | |
| Immunity 26 GHz to 1 GHz (option 18 GHz) |
Standard features:
| Single leaf RF door 914 x 2133 mm high | |
| Two waveguide vents 300 x 300 mm | |
| One connector panel 300 x 300 mm with two type N connectors | |
| One 30 amp two wire filter | |
| Two 5 amp filters for lighting | |
| Earth/ground stud | |
| Full or supervised installation |
Absorber Lining:
| Walls and ceiling covered with ferrite absorber panels. | |
| Patch of ferrite panels for placing on the floor between the EUT and antenna for immunity measurements. |
Typical Options:
| Hybrid absorber to increase frequency to 18 GHz (3.3 m wide
Model 1224 variant) is generally preferred for this application | |
| Low profile turntable or raised ground plane (reflective) floor and recessed turntable | |
| RF door 1200 mm wide | |
| Field uniformity measurement per EN 61000-4-3 | |
| Site attenuation measurement per EN50147-2 / ANSI C63.4 |
Typical Standards:
| EN 50147-part 2 | |
| ANSI C63.4 | |
| EN 61000-4-3 | |
| Various CISPR |
A larger version of the original Model 1023 Compact Anechoic Chamber (pictured above), this variant is designed for full compliant immunity measurements per EN 61000-4-3 and pre compliant emissions measurements to the procedures of EN50147 part 2 / ANSI C63.4. This larger compact chamber is generally partially lined with hybrid absorber for operation up to 18 GHz.
Nominal size:
|
7.2 long x 3.6 m wide x 3.6 m high |
Frequency range:
|
Emissions 30 MHz to 18 GHz (with hybrid absorber) | |
|
Immunity 26 GHz to 18 GHz (with hybrid absorber) |
Standard features:
|
Single leaf RF door 914 x 2133 mm high | |
|
Two waveguide vents 300 x 300 mm | |
|
One connector panel 300 x 300 mm with two type N connectors | |
|
One 30 amp two wire filter | |
|
Two 5 amp filters for lighting | |
|
Earth/ground stud | |
|
Full or supervised installation |
Absorber Lining:
|
Walls and ceiling covered with ferrite absorber panels | |
|
Patch of ferrite panels for placing on the floor between the EUT and antenna for immunity measurements | |
|
Hybrid absorber on EUT end wall | |
|
Hybrid absorber on critical areas of side walls and ceiling | |
Hybrid absorber over ferrite floor patch for testing above 1 GHz |
Typical Options:
|
Hybrid absorber included as standard for 18 GHz | |
|
Low profile turntable or raised ground plane (reflective) floor and recessed turntable | |
|
RF door 1.2 m wide | |
|
Field uniformity measurement per EN 61000-4-3 | |
|
Site attenuation measurement per EN50147-2 / ANSI C63.4 |
Typical Standards:
|
EN 50147-part 2 | |
|
ANSI C63.4 | |
|
EN 61000-4-3 | |
|
Various CISPR |